Stream Device Tips and Tricks

Rory Potter edited this page Oct 4, 2018 · 11 revisions

Wiki > The Backend System > IOCs > Stream Device Tips and Tricks

Waveforms

With any waveform record protocol, you need to specify a separator to split the input or output string. For example, to split on commas use separator = ","; within the protocol function for the record.

Reading in a string into a waveform of strings

If you have a waveform record of FTVL STRING, then you have to be careful about which formatter you use to read in that string.

Instead of using %s as your formatter, you need to use a character set, e.g. %[A-Z0-9a-z.+-], which does not include the separator you are splitting on. If you use %s, then the whole string will be read into the zeroth element of the array, and won't be split on the separator.

Example waveform of strings record and protocol file from Keithley 2001

Record

record(waveform, "$(P)SCAN:BUFF"){
    field(DESC, "Reads value from the buffer")
    field(DTYP, "stream")
    field(FTVL, "STRING")
    field(NELM, "10")
    field(INP, "@keithley_2001.proto read_buffer $(PORT)")
}

Protocol function

read_buffer {
    separator=",";
    out ":DATA:DATA?";
    in  "%[A-Za-z0-9.+-]";
}

Multi-value Protocols

Dealing with multi-value stream device protocols

It often happens that a single read query to a device returns multiple values, which we ultimately need to store in separate PVs. Similarly, sometimes a single write command requires multiple values. Below are some tips on how to deal with multi-value read protocols, but the same kind of tricks can be applied to write protocols too.

Passing the PV names as protocol parameters

Here is an example where a single status read query returns three different values, which correspond to three different PVs. The first PV triggers the protocol, and in doing so passes the name of the second and third PVs as arguments to the protocol:

## Read the status value 1. This also populates the status value 2 and 3.

record(ai, "$(P)VALUE1") {
  field(DESC, "Status value 1")
  field(DTYP, "stream")
  field(INP, "@ls336.proto getValues($(P)VALUE2,$(P)VALUE3) $(PORT)")
  field(SCAN, "$(SCAN) second")
}

## Read the status value 2.
record(bi, "$(P)VALUE2") {
  field(DESC, "Status value 2")
  field(ZNAM, "Off")
  field(ONAM, "On")
}

## Read the status value 3.
record(bi, "$(P)VALUE3") {
  field(DESC, "Status value 3")
  field(ZNAM, "OK")
  field(ONAM, "Error")
}

The corresponding protocol would be:

getValues {
   out "VALUES?";
   in "%f,%(\$1)d,%(\$2)d";
}

where \$1 and \$2 indicate the first and second argument, respectively. The protocol will redirect the second and third value to the PVs specified.

As our PV prefix $(P) tends to be quite long, often the INP field calling the protocol ends up being longer than the max number of allowed characters. We can shorten the INP field by passing the prefix as a separate argument (and remove blank spaces between arguments!):

...
field(INP, "@ls336.proto getValues($(P),VALUE2,VALUE3) $(PORT)")

and the prefix can be pre-pended to the PV names inside the protocol:

...
in "%f,%(\$1\$2)d,%(\$1\$3)d";}

If the protocol returns many values, even this approach may result in too long INP fields. One solution would be to pass the prefix $(P) as an argument and hard-code the rest of the PV names inside the protocol, but we try to avoid this as much as possible as it introduces extra coupling between the protocol and the db file. Read on for other tricks!

Reading into the inputs of a calc record

If you can't fit all the PV names in a single INP field, you can redirect them to the input fields of a calc record instead. We need a trigger PV which calls the protocol, and a buffer calc record to temporarily store the parsed values. The final PVs can then fetch their values from the calc record:


## Trigger the read protocol. Parsed values are stored in a buffer calc PV.
record(bi, "$(P)READ_DO") {
  field(DTYP, "stream")
  field(INP, "@ls336.proto getValues($(P)BUFFER) $(PORT)")
  field(SCAN, "$(SCAN) second")
}

## Store the parsed values
record(calc, "$(P)BUFFER") {
  field(CALC, "0")
}

## 
## Read the status value 1. 
##
record(ai, "$(P)VALUE1") {
  field(INP, "$(P)BUFFER.A CP")
  ...
}

## 
## Read the status value 2.
##
record(bi, "$(P)VALUE2") {
  field(INP, "$(P)BUFFER.B CP")
  ...
}

## 
## Read the status value 3.
##
record(bi, "$(P)VALUE3") {
  field(INP, "$(P)BUFFER.C CP")
  ...
}

and the protocol becomes:

...
in "%(\$1.A)f,%(\$1.B)d,%(\$1.C)d";

If you have to read strings as well as numbers, you can always use a scalcout record instead of a calc.

... and if everything fails

Another way is to read the entire input as a string into a string record, and then use various scalcout or asub records to parse the individual bits and pieces. An example of this usage can be found in the Linkam95 IOC.

Other considerations

The error on disconnection is not passed through from stream and so you may consider doing this via the error setter; see IOC Utilities for details.