TPGx00 is shorthand for the TPG300 and the TPG500 Pfeiffer Vaccuum Gauges which use the same TPG300
IOC, which controls a series of four pressure sensors (A1, A2, A3, A4).
To configure the IOC for the 500, define the MODEL macro: it defaults to TPG300
These devices are similar to the TPG26x and TPG36x, but have 4 pressure sensors to read from, and have a slightly more complicated IOC. All their manuals can be found in <share>\ISIS_Experiment_Controls\Manuals\Pfeiffer TPG<xxx>
.
The TPG500 is the slightly newer version of the TPG300; the main differences between these two devices is in their communication protocols. See the manuals for what these properties are used for.
300 | 500 | Affected commands | |
---|---|---|---|
Switching Functions | 1, 2, 3, 4, A, B |
1, 2, 3, 4 |
SPx |
Switching Function assignments | No Assignment, A1, A2, B1, B1, A1 self-monitoring, A2 self-monitoring, B1 self-monitoring, B1 self-monitoring |
Off, A1, A2, B1, B2, On |
SPx |
Units | mbar, Torr, Pa |
hPascal, mbar, Torr, Pa, Micron, Volt, Ampere |
UNI/UNIx |
Some less explicit differences are as follows:
- Communication terminators
- in the TPG300 manual, there is a mismatch of
<CR><LF>
and<CR>
receive acknowledgement terminators. The TPG500 consistently uses<CR><LF>
. This may be a typo in the TPG300 manual, but luckily as all the commands we use for the TPG300 are listed as using<CR><LF>
, this shouldn't cause any headaches.
- in the TPG300 manual, there is a mismatch of
-
SPS
command- Although the 300 and 500 have a mismatch of switching functions,
SPS
returns status values for all switching functions (e.g.1|2|3|4|A|B
) for both models. However, we don't actually care about the return values for SFsA
andB
for either model, so these last two values are skipped in the protocol file. Regardless, it's worth noting that for the 500 the values returned for A and B are:- 1 if an automatic function is active for sensor A1 (for SF
A
) or B1 (for SFB
) and the sensor is ON - 0 otherwise
- 1 if an automatic function is active for sensor A1 (for SF
- Although the 300 and 500 have a mismatch of switching functions,
-
SPx
command- This command returns a couple of parameters, one of which is the 'switching function assignment'. As noted above, these differ between the 300 and 500, but most importantly,
B1
being repeated twice isn't a typo - this is specified in the manual. By contrary, the 500 does have aB2
. - For the 500, this command returns an extra parameter which is the 'ON-Timer', but we don't care about this so is skipped in the protocol file.
- This command returns a couple of parameters, one of which is the 'switching function assignment'. As noted above, these differ between the 300 and 500, but most importantly,
The IOC currently supports the use following commands with the TPGx00 devices:
Command | Description |
---|---|
Pxx , where xx is A1|A2|A3|A4
|
Gets the pressure measurement from the specified channel |
ERR |
Get the error status of the device |
SPS |
Get the statuses of all switching functions 1|2|3|4|A|B
|
SPx , where x is 1|2|3|4 (+ A|B for 300) |
Get/set the threshold settings for a specified switching function |
UNI |
Get the unit of measurement |
UNIx , where x is 1|2|3 (+ 0|4|5|6 for 500) |
Set the current unit of measurement |
At the moment, there is some level of protection for using invalid commands between devices (e.g. sending a UNI6
to a TPG300):
-
$(P)UNITS:SP
(note: not$(P)UNITS
) goes into alarm if set outside the states defined for the current model -
$(P)FUNCTION
has protective PV$(P)FUNCTION:VALID
which only allows switching function SP/RB PVs to interact with the protocol file IFF$(P)FUNCTION
is in a 'valid' state.- For example, if you try to query the switching function
B
on a TPG500,SPB
cannot be sent to the TPG. However, as of #7458, no alarm is raised if$(P)FUNCTION
is set toB
, as is with:UNITS:SP
.
- For example, if you try to query the switching function